Strona 2

10 2.2 - Main Menu When you power the MC10X up, it will automatically boot and start the MC10X program on the controller. The unit will also home

Strona 3

11 Figure 2.3.1 Volume – Any integer value from 1 to 9999uL may be selected. Speed – This adjusts the percentage of the pumps maximum speed. More

Strona 4

12 Orientation– Select between the Portrait or Landscape plate orientation. This is determined by which manifold you are working with.12 cha

Strona 5 - Manifold port

13 Figure 2. 4.1 Volume – This is the total volume that will be primed through the system. Enter a value between 120 and 9999uL. Speed – This adj

Strona 6

14 Figure 2.5 1 2.6 - Auto-Prime Priming can be programmed to occur at a user specified interval. This allows the MC10X to dispense a certain a

Strona 7 - - Manifold

15 Figure 2.6.1 Time – This is the interval in seconds; a time of 0 deactivates this feature Volume – This is the total volume that will be moved

Strona 8 - Figure 1.4.1

16 Figure 2.7.1 Figure 2.7.2

Strona 9

17 Adjust Prime Position – This allows the user to jog the dispensing manifold to the position where they would like all priming functions to occur.

Strona 10 - 2.2 - Main Menu

18 Figure 2.7.4 Figure 2.7.5 Adjust Track/Nest Height – This value should be evaluated and adjusted if manifold types or plate nests are changed.

Strona 11

19 plate nest without touching it. To adjust, place a sheet of paper or business card on the plate nest. Jog the manifold down until it touches the p

Strona 12

2 Index Introduction ... 3 Section 1 - S

Strona 13

20 Figure 2.7.7 Adjust Row/Column Spacing - This setting is set by the manufacturer and should not be adjusted. # Of Tips in Manifold - This setti

Strona 14

21 Figure 2.8.1

Strona 15

22 Section 3 - Maintenance of the Mirco10X Fluid Path Note: Please turn power off and disconnect all cables from the rear of the Micro10x before

Strona 16

23 Figure 3.1 2 Figure 3.1.3 Pivot Bearing

Strona 17

24 3.2 - Disassembly of the Pump Head Module After the pump head module is removed from the pump you will able to access the Teflon adapters. The

Strona 18

25 Figure 3.2 3 Figure 3.2.4 With the Seal nut loose you will carefully slide the piston, seals and seal nut out of the cylinder case (Figure 3.2

Strona 19

26 Figure 3.2.5 3.3 – Cleaning of the Pump Assembly All of the pump parts may be cleaned in an ultrasonic cleaner with a dilute detergent solutio

Strona 20

27 Item # QTY Part Number Description 1 1 PMP-1018 Cylinder Nut 2 1 PMP-1021 Gland Washer, ¼”. Teflon 3 2 PMP-1019 Lip Seal, ¼”, Rulon RA 4 1 PMP-101

Strona 21 - Figure 2.8.1

28 Section 4 – Micro10x Validation Section 4.1 – General Validation After Cleaning To Ensure that the unit is dispensing properly after cleaning yo

Strona 22 - Pump head

29 Section 5 - Hardware Specifications Section 5.1 - General NOTE: Specifications are subject to change without notice. Micro10X:  Plate

Strona 23 - Pivot Bearing

3 Introduction The Hudson Robotic Micro10x™ is a fast, compact, accurate benchtop instrument for reliable, automated bulk reagent dispensing into al

Strona 24

30  Liner Bore Size: ¼”  Tubing Material: Peek fittings on Telfon tubing  Max Shot Size 225ul per revolution  Min Shot Size 12ul per rev

Strona 25

4 Section 1 - Setup 1.1 - Tubing The MC10X is shipped with two tube assemblies (Figure 1.1.1). The tube assembly with the two Flangeless fittings s

Strona 26

5 Figure 1.1.2 Figure 1.1.3 Right pump port Manifold port Left pump port

Strona 27

6 The priming trough uses a short silicon drain tube and waste basin. The silicon tube can be replaced with a longer one if needed. Standard silico

Strona 28

7 1.3 - Manifold Install the manifold with the inlet port closest to the pump head. Use the provided thumbscrews to secure the manifold ont

Strona 29

8 Figure 1.4.1 Controller Cable DC Inlet for Power cord Pump Cable Pump Cable Port for optional 4-way valve Serial Port Power Switch

Strona 30 - Internal T4.0 A/250 V

9 Section 2 - Operation of Micro10X through the Touch Screen Controller 2.1 - Controller Features The Micro10X comes with a windows based touch sc